Repository navigation
USB and Remote Mode
Caution
The USB port is not isolated from the measurement inputs. While the device is connected to a computer, the probe ground and the inputs are tied to the computer's USB ground. Measuring a circuit with another ground potential, such as mains-powered equipment, a car or anything not floating, can short that potential through the device and the computer: it can damage both and is dangerous. For such measurements, unplug the USB cable and run on battery, or use a USB isolator. Never measure mains voltages with the USB cable connected.
In the running firmware the device appears on the computer as a USB instrument with two interfaces:
| Interface | Use |
|---|---|
| USBTMC | SCPI commands and binary waveforms (VISA-based software) |
| CDC (virtual serial port) | The same SCPI commands as text |
Current operating systems need no driver for the serial port; USBTMC uses your VISA library. See the SCPI Reference.
- USB data works only in the 200/100 profile.
- Connect directly to a computer port. Disconnects are known, especially through USB hubs.
- USBTMC accepts large read limits from host software such as ngscopeclient on Linux. It returns the available reply and ends the transfer when the reply is complete.
- For files, open Recovery: it shows the storage as a USB drive.
The instrument identity follows the active screen:
- Oscilloscope profile while the oscilloscope is in use;
- Multimeter profile while the multimeter is in use.
Both profiles use command sets compatible with common PC software for bench oscilloscopes and multimeters. Generator commands work in both. Switching between oscilloscope and multimeter re-connects USB (about 0.3 s): reopen the connection and check *IDN?. Switching to the generator screen does not change the profile.
Settings → Files & USB → USB Remote or SYST:REM ON.
- The computer has control; the local keys are locked.
- The screen stays on and keeps showing live data.
- Entering Remote switches to the 200/100 profile if needed.
-
To leave: hold MENU (the notice reads USB Remote off), or send
SYST:REM OFF. Holding REL/Power still switches the device off. - After power-on the device is always in local (standalone) mode.
SYST:REC closes all files, stops USB and restarts into Recovery, which then shows the storage as a USB drive.